ZR-3922 Ambient Air Sampler

Short Description:

 Ambient Air Sampler is a portable instrument. It uses filter membrane to capture particles in the ambient air (TSP, PM10, PM2.5). The solution absorption method is used to collect various harmful gases in the ambient atmosphere and indoor air. It can be used for aerosol monitoring by environmental protection, health, labor, safety supervision, scientific research, education and other departments.


  • Ambient air sampling flow: (15~130)L/min
  • Atmospheric sampling flow: (0.1~1.5)L/min
  • Weight: About 4.0kg(battery include)
  • Power consumption: ≤120W
  • Size: (length 310×width 148×height 220)mm
  • Product Detail

    Specification

    Standards

    HJ 93-2013  Technical requirements and detection methods of ambient air particulate matter (PM10 and PM2.5) sampler

     HJ 618-2011  Ambient air PM10 and PM2 5 gravimetric method

     HJ 656-2013  Technical specification for manual monitoring method (gravimetric method) of ambient air particulate matter (PM 2.5)

     HJ/T 374-2007   Technical requirements and detection methods of total suspended particulate matter sampler

     HJ/T 375-2007  Technical requirements and testing methods of ambient air sampler

     JJG 943-2011   Verification regulation of total suspended particulate matter sampler

     JJG 956-2013   Verification regulation of atmospheric sampler

     Q/0214 ZRB010-2017  Comprehensive sampler for ambient air particulate matter

    Features

     4.3-inch color screen ,touch operation and the operation is simple

     Compact size, light in weight, easy to carry

      Built-in Lithium-Polymer Rechargeable Battery

      Three channels of simultaneous sampling can be used to collect particulate matter and gaseous pollutants in the air

      Rain proof, dust-proof, anti-static and anti-collision design can ensure normal operation under the conditions of rain, snow, dust and heavy haze

       Various sampling methods, such as constant sampling time, continuous sampling time and 24-hour sampling, can be realized

      The cutter(TSP / PM 10 / PM 2.5) is made of aluminum alloy with anti-static adsorption

      Power-off memory function,continue to sample procedure when recovery

      Support data storage and export data with USB

      Print with wireless Bluetooth

    Deliver Goods

    deliver goods Italy
  • Previous:
  • Next:

  • Parameter Range Resolution Error
    Ambient air sampling flow (15~130)L/min 0.1L/min ±5.0%
    Ambient air sampling time 1min~99h59min 1s ±0.1%
    Load capacity When the flow is 100L / min, the load capacity is > 6kpa
    Atmospheric sampling flow (0.1~1.5)L/min 0.01L/min ±2.0%
    Atmospheric sampling time 1min~99h59min 1s ±0.1%
    Ambient atmospheric pressure (60~130)kPa 0.01kPa ±0.5kPa
    Temperature range of incubator ≥15℃ 0.1℃ ±2℃
    temperature (-30~50)℃
    Noise <65dB(A)
    Discharge duration Three circuits work at the same time, TSP load is 2KPa, and the discharge time is > 6h
    Charging time Internal charging < 12h, external charging < 4H
    Power Supply AC(220±22)V,(50±1)Hz
    Size (length 310×width 148×height 220)mm
    Weight About 4.0kg(battery include)
    Power consumption ≤120W
    Write your message here and send it to us