ZR-6012 Aerosol photometer

Short Description:

For leakage detection for HEPA Filter, it is well known to use aerosol photometer for testing. ZR-6012 Aerosol photometer is a professional testing equipment which is used to test whether there is leakage on HEPA filter.

The downstream of aerosol concentration can be extended to 0.0001 μg/L, and upstream can be extended to 700μg/L.


  • Sampling flowrate: 28.3L/min
  • Concentration detection range: 0.01~125μg/L;Ultra-low concentration can be extended to 0.0001 μg/L; Ultra-high concentration can be extended up to 700 μg/L
  • Leakage detection: 0.0001%~100%
  • Data storage: 100000 groups
  • Size: (length 300×width 330×height 184)mm
  • Weight: About 8.9kg(Battery included) /About 15kg(Outer packing, instrument, accessories, etc.)
  • Power consumption: <150W
  • Product Detail

    Applications

    Specification

    Accessory

    ZR-6012 Aerosol photometer is used to test whether there is leakage on HEPA filter. According to the principle of light scattering, it is portable, yet rugged for in-situ filtration system integrity testing.

    The instrument conforms to NSF49 / IEST / ISO14644-3, could realize rapid detection of upstream and downstream concentration detection and real-time display leakage on the host and handheld device, and could find leaking position rapidly and accurately.

    Standards

    NSF/ANSI 49-2019  Biosafety cabinetry

    ISO14644-3:2005    Cleanrooms and associated controlled environments—Part 3:Test methods

    GB 50073-2013       Code for design of clean workshop

    GB 50591-2010       Code for construction and acceptance of cleanroom

    2010 GMP for pharmaceutical products-plant and equipment

    YY0569-2005          Biosafety cabinet

    JJF 1800-2020       Calibration specification for aerosol photometer

    JJF 1815-2020       Calibration specification for class II biosafety cabinet

    Features

    • Powerful function

    >  Stable digital photometer.

    >  Dynamic Range: 0.0001μg/L~700μg/L.

    • Data query

    >  USB & Printer available for real-time reporting.

    >  Audit Tracing & assign permissions, further guarantee data integrity.

    >  Sampling data can be imported to PC.

    • Automatic reminder

    >  When exceed the set value, light & voice alarm.

    >  Self protection in case of failure.

    • Good human-computer interaction

    >  5.0 -inch color screen, touch operation.

    >  Lightweight, portable and equipped with suitcase, easy to carry.

    >  Built-in battery(optional)≥ 3.5H.

    Deliver Goods

    deliver goods Italy
  • Previous:
  • Next:

  • Medical facilities and cleanrooms                                                                                                                

    Aerosol photometer (2)

    Biosafety cabinets and fume hoods

    Aerosol photometer (3)

    Independent filter certifiers    

    Aerosol photometer (1)

    Pharmaceutical manufacturers

    Aerosol photometer (4)

     

     

    Main parameters

    Scope

    Sampling flow

    28.3L/min,±2.5%

    Concentration detection range

    0.01~125μg/L;Ultra-low concentration can be extended to 0.0001 μg/L; Ultra-high concentration can be extended up to 700 μg/L

    Detection of leakage rate

    0.0001%~100%

    Detection precision

    1% of the value in the range of 0.01% to 100%

    Detection repeatability

    0.5% of the value in the range of 0.01% to 100%

    Data storage capability

    100000 groups

    Size(W×D×H)

    (300×330×184)mm

    Weight

    8.9 kg

    Packed weight

    15kg(Outer packing, instrument, accessories, etc.)

    Power

    AC110V/60Hz

    Power consumption

    <150W

    Environmental temperature

    10℃~35℃

    Environmental humidity

    5%~85%RH(No dew, no ice)

    Storage requirements

    -10℃~ 40℃, RH<85%, no dew

    Test medium

    PAO, DOP and other aerosol types

    Noise

    <65dB(A)

    When detecting the leakage of high-efficiency filter, you need to cooperate with aerosol generator. It emits aerosol particles with different sizes, and adjusts the aerosol concentration as required to make the upstream concentration reach 10 ~ 20ug / ml. Then the aerosol photometer will detect and display the concentration of the particle mass.

    Aerosol photometer (1)

    ZR-1300A  Aerosol Generator

    Write your message here and send it to us